The DH-2000-BAL uses innovative filtering technology to produces a smooth spectrum across entire UV-NIR range. This technology also eliminates the alpha deuterium line in the visible region. Using a combination of deuterium and halogen lamps, the DH-2000 is ideal for measuring a sample that has multiple features in different spectral regions or for analyzing a variety of different samples in your lab.

All deuterium sources have an alpha line, a sharp spectral feature centered at 655 nm. This feature and other deuterium lines produce “unbalanced” output in the deuterium and halogen sources. Simply adjusting the integration time of the spectrometer to suppress the alpha line does not solve the problem as the efficiency of the system at UV wavelengths drops significantly, compromising signal-to-noise performance.

Product Overview

  • Innovative filtering -- eliminates D-alpha line in deuterium source
  • Balanced output from 210-2500 nm -- eliminates problems associated with saturation
  • Powerful and stable -- deuterium (25 W) and halogen lamps (20 W) combined in a single light source
  • Integrated shutter -- helps protect light-sensitive samples
  • Easy-to-replace, pre-aligned bulbs -- minimal downtime needed for bulb replacement

Specifications

Select model to see full specifications

Engineering:

Wavelength Range:
210nm - 2.5μm
Source:
Deuterium & Tungsten Halogen
Nominal Bulb Power:
25 W (deuterium)
20 W (tungsten halogen)
Typical Output Power:
194 µW (deuterium bulb)
615 µW (tungsten bulb)
Source Lifetime:
1000 hours
Stability of Light Source Output:
≤0.1%/hour @ 254 nm (deuterium)
≤0.1%/hour @ 700 nm (tungsten halogen)
Warm-up Time:
25 minutes

Environmental:

Safety & Regulatory:
CE
ROHS
WEEE
Operating Humidity Specification:
5-95% without condensation at 40 °C
Operating Temperature:
5 °C – 35 °C

Physical:

Shutter:
Yes
Fiber Connector:
SMA 905
Dimensions:
15 x 13.5 x 28.5 cm
Integrated Filter Holder:
No
Power Consumption:
Approximately 78VA
Trigger/Shutter Input Connection:
SUB-D-15 pin
Trigger/Shutter Input Signal:
TTL
Up to 2.5 Hz maximum
Weight:
5.5 kg
Integrated Cuvette Holder:
No
Power Requirements:
85-264 V 50/60 Hz