QR200-REF-UV-VIS

Part Number: QR200-REF-UV-VIS

Lead Time: 2 weeks

Total Price

Online prices exclude taxes, logistics and customs and trade tariffs. Select your “ship to” country for accurate pricing. All offline purchases require a quotation.

Couldn't update

This product is not available to purchase online.
Please contact us for further details.

Product Details

A reflection/backscattering probe with a built-in reference leg is a convenient solution for long-term measurements and experiments in which variation or instability in the spectral output of the light source is expected to affect the reflection spectrum.

Available in both premium- and lab-grade, these probes include a reference leg that splits off from the light source connector for monitoring your illumination (or reference) source via a second spectrometer.

Product Overview

  • Probe design -- 6-around-1 fiber bundle design at probe end for sample measurements
  • Setup -- 7-fiber leg connects to light source, single-fiber leg connects to spectrometer for best performance
  • Reference leg -- one optical fiber splits off from the light source connector for reference monitoring (6-around-1 design)
  • Probe type -- probe comprises UV-Visible (300-1100 nm) fibers

Reflection-Probe-Ref-Leg_NEW.jpg

Specifications

Select model to see full specifications

Engineering:

Fiber Core Size:
200 µm
Length:
2 m
Jacket:
Silicone-coated steel monocoil
Probe Ferrule Diameter:
6.35 mm (1/4″)
Breakout:
Midway point of assembly at 1 meter
Buffer Materials:
Polyimide
Cladding OD:
10 µm
Cladding OD Tolerance:
± 4 µm
Connector:
SMA 905
Fiber Core Size Tolerance:
± 4 μm
Long Term Bend Radius:
8 cm
Maximum OD:
243 µm
Primary Buffer OD:
10 µm
Primary Buffer OD Tolerance:
± 5 µm
Probe Fiber Bundle:
6 illumination fibers around 1 read fiber
Short Term Bend Radius:
4 cm
Probe Ferrule Length:
76.2 mm (3″)
Probe Ferrule Material:
Stainless steel