Details
A reflection/backscatter probe is a compact, versatile sampling option to measure diffuse and specular reflectance, backscatter, or fluorescence from samples as varied as solids, solutions, and powders. Reflection and backscatter measurements can provide quantitative information about the color, appearance, and chemical composition of a sample.
Overview
- General purpose -- ideal for diffuse or specular reflectance from solids; fluorescence; color
- XSR fibers -- probe assembly comprises extreme solarization-resistant fibers (180-800 nm)
- Probe design -- 6-around-1 fiber bundle design, with 6-fiber leg connecting to light source and single-fiber leg connecting to spectrometer
Specifications
Subject to change depending on model type
Engineering:
- Fiber Core Size:
- 230 µm
- Jacket:
- Silicone-coated steel monocoil
- Length:
- 2 m
- Probe Ferrule Diameter:
- 6.35 mm (1/4″)
- Wavelength Range:
- 180nm - 800nm
- Breakout:
- Midway point of assembly at 1 meter
- Buffer Materials:
- Aluminum, Polymer
- Cladding OD:
- 250 µm
- Cladding OD Tolerance:
- ± 13 µm
- Connector:
- SMA 905
- Fiber Core Size Tolerance:
- ± 12 μm
- Long Term Bend Radius:
- 4 cm
- Maximum OD:
- 380 µm
- Maximum OD Tolerance:
- ± 20 µm
- Numerical Aperture:
- 0.22 ± 0.02 (equivalent to an acceptance angle of 24.8° in air)
- Primary Buffer OD:
- 300 µm
- Probe Ferrule Length:
- 76.2 mm (3″)
- Probe Ferrule Material:
- Stainless steel
- Probe Fiber Bundle:
- 6 illumination fibers around 1 read fiber
- Short Term Bend Radius:
- 2 cm
Environmental:
- Operating Temperature:
- -50 to 80 °C