Custom Configured Flame Spectrometer
Details
Flame series spectrometers deliver all the benefits of miniature, modular spectroscopy — without comprise.
These compact spectrometers integrate performance-enhancing features and are produced using industry-leading manufacturing techniques. Enjoy the benefits of greater thermal stability and lesser unit-to-unit variability, plus the freedom of interchangeable slits, simple device connectors and LED status indicators.
Overview
- Modular — configurable to your application within the wavelength range 200-1000 nm
- Compatible — works with Ocean Insight light sources, accessories and software
- Easy to use — plug and play via the micro USB connection
- User interchangeable slit — vary the resolution and throughput of the spectrometer on demand
- Indicator LEDs — indicator of power and data transfer status at all times
- Compact and lightweight — 89.1 mm x 63.3 mm x 34.4 mm and just 265 g
Take advantage of our technical guidance and this interactive spectrometer range-resolution tool.
Key Feature | Best For | Example Applications |
User Interchangeable Slits | Users who wish to vary resolution and throughput during measurements, or switch from absorbance to fluorescence in minutes | Life science and other labs using a variety of biological samples |
Indicator LEDs | Convenient visual reference for spectrometer operation and status | Teaching and general lab use |
Thermal Stability | Applications that require repeatable results in industrial and other environments with varying temperatures and conditions | LED binning and light metrology, process monitoring |
Reduced Unit to Unit Variation | OEM applications such as uncalibrated sensitivity and other measurement needs where users benefit from low variance | OEM manufacturers of analytical instrumentation |
Configurability | Optimizing your spectrometer for application-specific requirements; adjusting range, throughput and resolution and adding features such as filters as required | Laser characterization, low signal fluorescence and optimizing for specific absorbance bands |
Plug & Play Operation | Users that want the convenience of simple, fast operation via the microUSB connection; enables users to take the measurement to the sample | Remote sensing measurements in the field, including air and water quality monitoring and solar irradiance |
Ease of Integration | OEMs and developers who need to integrate a spectrometer as part of a system via microUSB or RS-232 | Engineering labs, developers, OEM manufacturers; works with LabVIEW and other design platforms |
Specifications | FLAME-S | FLAME-T |
SPECTROSCOPIC | ||
Optical resolution: | ~0.1-10.0 nm FWHM (configuration dependent) | |
Signal-to-noise ratio: | 250:1 (full signal) | 300:1 (at full signal) |
Dark noise: | 50 RMS counts | |
Dynamic range: | 1300:1 for a single acquisition, 8.5 x 10^7 (system) | |
Integration time: | 1 ms – 65 seconds | 3.8 ms – 10 seconds |
Corrected linearity: | >99% | |
ELECTRONICS | ||
A/D resolution: | 16 bit | |
Power consumption: | 250 mA @ 5 VDC | |
Inputs/Outputs: | 8 x user programmable GPIOs | |
Trigger modes: | 4 modes | |
Strobe functions: | Yes | |
Gated delay feature: | Yes | |
Connectors: | USB, high density 40 pin JAE DD4 connector | |
DETECTOR | ||
Detector: | Sony ILX511B linear silicon CCD array | Toshiba TCD1304AP linear silicon CCD array |
Detector range: | 190-1100 nm | 190-1100 nm |
Pixels: | 2048 | 3648 |
PHYSICAL | ||
Dimensions: | 89.1 mm x 63.3 mm x 31.9 mm (34.4 mm with feet) | |
Weight: | 265 g |