Flame series spectrometers deliver all the benefits of miniature, modular spectroscopy — without comprise.

These compact spectrometers integrate performance-enhancing features and are produced using industry-leading manufacturing techniques. Enjoy the benefits of greater thermal stability and lesser unit-to-unit variability, plus the freedom of interchangeable slits, simple device connectors and LED status indicators.

Overview

  • Modular — configurable to your application within the wavelength range 200-1000 nm
  • Compatible — works with Ocean Insight light sources, accessories and software
  • Easy to use — plug and play via the micro USB connection
  • User interchangeable slit — vary the resolution and throughput of the spectrometer on demand
  • Indicator LEDs — indicator of power and data transfer status at all times
  • Compact and lightweight — 89.1 mm x 63.3 mm x 34.4 mm and just 265 g
Key Feature Best For Example Applications
User Interchangeable Slits Users who wish to vary resolution and throughput during measurements, or switch from absorbance to fluorescence in minutes Life science and other labs using a variety of biological samples
Indicator LEDs Convenient visual reference for spectrometer operation and status                         Teaching and general lab use
Thermal Stability Applications that require repeatable results in industrial and other environments with varying temperatures and conditions LED binning and light metrology, process monitoring
Reduced Unit to Unit Variation OEM applications such as uncalibrated sensitivity and other measurement needs where users benefit from low variance OEM manufacturers of analytical instrumentation
Configurability Optimizing your spectrometer for application-specific requirements; adjusting range, throughput and resolution and adding features such as filters as required Laser characterization, low signal fluorescence and optimizing for specific absorbance bands
Plug & Play Operation Users that want the convenience of simple, fast operation via the microUSB connection; enables users to take the measurement to the sample Remote sensing measurements in the field, including air and water quality monitoring and solar irradiance
Ease of Integration OEMs and developers who need to integrate a spectrometer as part of a system via microUSB or RS-232 Engineering labs, developers, OEM manufacturers; works with LabVIEW and other design platforms
Specifications FLAME-S FLAME-T
SPECTROSCOPIC
Optical resolution: ~0.1-10.0 nm FWHM (configuration dependent)
Signal-to-noise ratio: 250:1 (full signal) 300:1 (at full signal)
Dark noise: 50 RMS counts
Dynamic range: 1300:1 for a single acquisition, 8.5 x 10^7 (system)
Integration time: 1 ms – 65 seconds  3.8 ms – 10 seconds
Corrected linearity: >99%
ELECTRONICS
A/D resolution: 16 bit
Power consumption: 250 mA @ 5 VDC
Inputs/Outputs: 8 x  user programmable GPIOs
Trigger modes: 4 modes
Strobe functions: Yes
Gated delay feature: Yes
Connectors: USB, high density 40 pin JAE DD4 connector
DETECTOR
Detector: Sony ILX511B linear silicon CCD array Toshiba TCD1304AP linear silicon CCD array
Detector range: 190-1100 nm 190-1100 nm
Pixels: 2048 3648
PHYSICAL
Dimensions: 89.1 mm x 63.3 mm x 31.9 mm (34.4 mm with feet)
Weight: 265 g